qPCR reaction setup & cycling protocols
Reaction setup, kit components and cycling conditions for every assay format we supply. Assay-specific values are listed on each product page.
One-step RT-PCR assay (RNA targets)
Reaction setup and cycling for primerprobe ready-to-mix assays on RNA targets using a one-step RT-qPCR master mix.
Reaction setup (20 uL)
| Component | Volume |
|---|---|
| 2X One-step RT-PCR Master Mix | 10.0 uL |
| RT Enzyme Mix | 0.5 uL |
| Primer-Probe Mix | 1.0 uL |
| Template RNA | 5.0 uL |
| Nuclease free water | 3.5 uL |
| Final volume | 20.0 uL |
Temperature profile
| Step | Temperature | Time | Cycles |
|---|---|---|---|
| Reverse transcription | 50 C | 15 min | 1 |
| Initial denaturation | 95 C | 2 min | 1 |
| Denaturation | 95 C | 15 sec | 40 |
| Annealing / extension (data acquisition) | 60 C | 45 sec | 40 |
Notes
- Keep the RT enzyme on ice and add it last.
- Include a no-template control and the supplied internal control in every run.
- Thaw primer-probe mixes in the dark, vortex and spin down before use.
For research use only. Not for use in diagnostic procedures.